DAVITEQ CHLORINE Cl2 GAS SENSOR, 20 PPM, SERI-4 SENSOR SPECIFICATION MEASURING SPECIFICATION Sensor type Daviteq Seri-4 electrochemical gas sensor Nominal range 0-20 ppm Maximum overload 200 ppm Resolution 0.01 ppm T80 response time < 60 s Typical Baseline Range (pure air, 20°C) -0.1 ppm .. +0.2 ppm Repeatability < 2% of reading or 0.05ppm whichever greater Accuracy +/- 5% of reading Zero stability 0 .. 0.05 ppm in pure Air, 20oC Temperature compensation Yes WORKING CONDITIONS Relative humidity range 15 % to 90 % RH non-condensing Temperature range -40 °C to 50 °C Pressure range Atmospheric ± 10% Humidity effect None SENSOR LIFETIME Expected Operation Life 2 years in the air Expected Long-Term Output Drift in air < 2 % of reading loss per month CROSS SENSITIVITY DATA The table below does not claim to be complete. Interfering gases should not be used for calibration. Interfering Gas Concentration (ppm) Reading (ppm) Br2 10 2.5 C2H4 100 0 CH2O 7 0 ClO2 3 ~3 CO 300 0 Ethanol (C2H5OH) 60 0 F2 8 ~8 H2 100 0 H2S 20 < -20 HCl 20 0 NH3 80 0 NO 50 0 NO2 20 ~20 O3 1 ~1 SO2 5 0