DAVITEQ FORMALDEHYDE GAS SENSOR CH2O, 10 PPM, SERI-4 SENSOR SPECIFICATION MEASURING SPECIFICATION Sensor type Daviteq Seri-4 electro-chemical gas sensor Nominal range 0-10 ppm Maximum overload 30 ppm Filter - Resolution 0.01 ppm T90 response time < 30 s Typical Baseline Range (pure air, 20°C) -0.1 ppm .. + 0.1 ppm Repeatability < 2% of reading Accuracy < +/- 5% of reading WORKING CONDITIONS Relative humidity range 15 % to 90 % RH non-condensing** Abrupt changes in RH causes a short term transient signal Temperature range -40 °C to 50 °C Pressure range Atmospheric ± 10% SENSOR LIFETIME Expected Operation Life 3 years in the air Expected Long-Term Output Drift in air < 2 % of reading loss per month We have other ranges: 50 ppm, and 1000 ppm. CROSS SENSITIVITY DATA Interfering Gas Concentration (ppm) Reading (ppm) CO2 100 < 20 Ethanol (C2H5OH) 30 1 H2 100 5 H2S 20 ~ 20 Methanol (CH3OH) 80 10 Methyl Mercaptan (MM, CH3SH) 10 ~6 Tert-Butyl Mercaptan (TBM, (CH3)3CSH) 10 ~6