DAVITEQ NITROGEN DIOXIDE NO2 GAS SENSOR, 20 PPM, SERI-4 SENSOR SPECIFICATION MEASURING SPECIFICATION Sensor type Daviteq Seri-4 electro-chemical gas sensor Nominal range 0-20 ppm Maximum overload 40 ppm Filter Inboard filter to remove acidic gases Resolution 0.1 ppm T90 response time < 40 s Typical Baseline Range (pure air, 20°C) -0.1 ppm .. + 0.2 ppm Repeatability < 2% of reading Accuracy +/- 5% of reading Zero Stability 0 .. 0.2 ppm in pure Air, 20oC Temperature compensation Yes WORKING CONDITIONS Relative humidity range 15 % to 90 % RH non-condensing Temperature range -40 °C to 50 °C Pressure range Atmospheric ± 10% SENSOR LIFETIME Expected Operation Life 2 years in the air Expected Long Term Output Drift in air < 2 % of reading loss per month CROSS SENSITIVITY DATA Interfering Gas Concentration (ppm) Reading (ppm) C2H4 100 0 CH2O 7 0 Cl2 20 15 CO 100 0 ETHANOL (C2H5OH) 60 0 H2 100 0 H2S 20 < - 20 HCL 20 0 NH3 80 0 NO 50 0 O3 1 1 SO2 5 0