HIGH PERFORMANCE GAS MEASUREMENT MODULE MAX-O3 Item code: MAX-O3-C5 Description: Internal Sensor module for measuring Ozone gas. Used for: MAQM-MAX SENSOR SPECIFICATION MEASURING SPECIFICATION Measuring technology Electro-chemical Measurement range 0-5 ppm Overload 50 ppm Resolution 0.001 ppm Lower Detectable Limit 0.001 ppm Accuracy +/- 2% of Reading value Linearity < 0.1% of Full scale Response Time (T80) < 60s Operating Pressure Atmospheric ± 10% Working Temperature -40 .. + 50 oC Working Humidity 15..90% RH non-condensing, continuous Sensor Life Time 2 years in clean air Drift in Air < 2 % signal loss per month Power supply 9..36VDC, max 30mA CROSS SENSITIVITY DATA The table below does not claim to be complete. Interfering gases should not be used for calibration. Important Application Notes: NO readily forms NO2 in the presence of O2.