DAVITEQ TRACE OXYGEN O2 GAS SENSOR 0-1%, SERI-4
KEY FEATURES
• Fast equilibration time
• Long-Life sensor with no consumable parts and lead-free (RoHS compliant)
APPLICATIONS
• Trace oxygen measurements
• Continuous Air Quality Monitoring
• Safety and Environmental Control
SENSOR SPECIFICATION
| MEASURING SPECIFICATION | |
| Sensor type | Daviteq Seri-4 electrochemical gas sensor |
| Nominal range | 0-1% |
| Maximum overload | 21% |
| Filter | come with the main device |
| Resolution | 0.01% |
| T90 response time | < 30s (via filter) |
|
Typical Baseline Range (20°C) |
< 0.01% in pure N2 |
| Repeatability | < 2% of reading |
| Accuracy | +/- 5% of reading |
| Zero Stability | 0 .. 0.01 % in pure Nitrogen, 20oC |
| Temperature compensation | Yes |
| WORKING CONDITIONS | |
| Relative humidity range |
15 % to 90 % RH non-condensing |
| Temperature range |
-40 °C to 50 °C |
| Pressure range |
Atmospheric ± 10% |
| SENSOR LIFETIME | |
|
Expected Operation Life |
3 years in the air |
|
Expected Long-Term Output Drift in air |
< 1.33 % of reading loss per year |
No Comments