HIGH PERFORMANCE GAS MEASUREMENT MODULE MAX-SO2

Item code: MAX-SO2-C20

Description: Internal Sensor module for measuring SO2 gas.

Used for: MAQM-MAX

SENSOR SPECIFICATION
MEASURING SPECIFICATION
Measuring technology Electro-chemical
Measurement range 0-20 ppm
Overload 100 ppm
Resolution 0.001 ppm
Lower Detectable Limit 0.009 ppm
Accuracy +/- 2% of Reading value

Linearity

< 0.1% of Full scale

Response Time (T90) < 15s
Operating Pressure Atmospheric ± 10%
Working Temperature -20 .. + 50 oC
Working Humidity 15..90% RH non-condensing, continuous
Sensor Life Time 2 years in clean air
Drift in Air < 2 % signal loss per month
Power supply 9..36VDC, max 30mA
CROSS SENSITIVITY DATA

The table below does not claim to be complete. Interfering gases should not be used for calibration.

image-1672111754713.png

Important Application Notes: