HIGH PERFORMANCE GAS MEASUREMENT MODULE MAX-O3

Item code: MAX-O3-C5

Description: Internal Sensor module for measuring Ozone gas.

Used for: MAQM-MAX

SENSOR SPECIFICATION
MEASURING SPECIFICATION
Measuring technology Electro-chemical
Measurement range 0-5 ppm
Overload 50 ppm
Resolution 0.001 ppm
Lower Detectable Limit 0.001 ppm
Accuracy +/- 2% of Reading value

Linearity

< 0.1% of Full scale

Response Time (T80) < 60s
Operating Pressure Atmospheric ± 10%
Working Temperature -40 .. + 50 oC
Working Humidity 15..90% RH non-condensing, continuous
Sensor Life Time 2 years in clean air
Drift in Air < 2 % signal loss per month
Power supply 9..36VDC, max 30mA
CROSS SENSITIVITY DATA

The table below does not claim to be complete. Interfering gases should not be used for calibration.

image-1672111110408.png

Important Application Notes: NO readily forms NO2 in the presence of O2.